ReBeatICG: Real-time Low-Complexity Beat-to-beat Impedance Cardiogram Delineation Algorithm

4 May 2021  ·  Una Pale, Nathan Müller, Adriana Arza, David Atienza ·

This work presents ReBeatICG, a real-time, low-complexity beat-to-beat impedance cardiography (ICG) delineation algorithm that allows hemodynamic parameters monitoring. The proposed procedure relies only on the ICG signal compared to most algorithms found in the literature that rely on synchronous electrocardiogram signal (ECG) recordings. ReBeatICG was designed with implementation on an ultra-low-power microcontroller (MCU) in mind. The detection accuracy of the developed algorithm is tested against points manually labeled by cardiologists. It achieves a detection Gmean accuracy of 94.9%, 98.6%, 90.3%, and 84.3% for the B, C, X, and O points, respectively. Furthermore, several hemodynamic parameters were calculated based on annotated characteristic points and compared with values generated from the cardiologists' annotations. ReBeatICG achieved mean error rates of 0.11ms, 9.72ms, 8.32ms, and 3.97% for HR, LVET, IVRT, and relative C-point amplitude, respectively.

PDF Abstract
No code implementations yet. Submit your code now

Tasks


Datasets


  Add Datasets introduced or used in this paper

Results from the Paper


  Submit results from this paper to get state-of-the-art GitHub badges and help the community compare results to other papers.

Methods


No methods listed for this paper. Add relevant methods here